driving torque (insertion/removal) of metallic bone screws

Stipulated the torque required to measure when a metal bone screw is screwed in and out of the standard material to confirm the consistency of the product under test. The constant force is maintained at 11.17N, at a rate of 3r/min, first screwed in 4 turns, then 4 turns out.


Test procedure:

Bone screws are inserted into and removed from a test block (may be predrilled); specifications to ASTM F1839; clamping via specimen grips.

Screw bits are mounted via a quick-acting chuck and the screw is driven at a test speed of 1 to 5 rpm in accordance with the standard. If required, a constant axial preload can be applied.

A minimum of four screw revolutions are required.


Result: 

Insertion/removal torque versus angle for the entire duration of the test

Axial load force versus travel for entire duration of test

Insertion Depth




Torque measuring range

0-5N.m

Resolution

0.001N.m

Torque   sensor accuracy

Less   than 0.5%F.S.

Speed   accuracy

±0.01mm/min

Axial   force load cell

200N/1000N   (fulfill axial pull-out test)

Load   cell accuracy

Less   than 0.5%F.S.

Load   cell resolution

0.001N

Linear   displacement sensor range

More   than 50m

Displacement   sensor accuracy level

Less   than 0.5%F.S.

Displacement   sensor resolution

0.01mm

Torsion   angle sensor

99999   degrees

Torsional   angle accuracy

±0.2   degree

Torsion   speed

1r/min   to 30r/min free adjusting

Effective   travel

400±10mm   (can be increased on demand)

The   coaxiality of the two chucks

  0.2mm

Standard   Power

220V,   50HZ, 1 phase, 0.4Kw

Working   system

MS   Win10


    