EN 728, ISO 11357 Differential scanning Calorimeter, Oxidative Induction Time testing machine

Oxidation induction time or OIT is a standardized test performed in a DSC ( Differential scanning Calorimeter) which measures the level of thermal stabilization of the material tested. The time between melting and the onset of decomposition in isothermal conditions is measured. The atmosphere is nitrogen up to melting and then oxygen. The typical temperature is 190-220 °C.


Oit.png


UT3335 Differential scanning calorimeter is one kind of thermal analysis method instrument, under controlled temperature, check the relation of temperature with the rate differential input in test sample and reference articles.  The curve recorded is DSC curve, heat flow rate dH/dt is Y-axis, temperature T or time t is X-axis, can measure various kind of thermodynamics and kinetic parameter, like specific heat capacity, reaction heat, heat of transformation, phase diagram, reaction rate, cystallization rate, sample purity etc.,  with wide temperature range (-100~800℃), high resolutions, less sample using. Used to test inorganic material, organic compounds, pharmaceutical analysis etc.,  can be used to do Oxidative Induction Time test, heat stability, glass transition temperature, melting point, cold crystallization, crystallization, phase transition,  etc.,         

 


Conform to standard EN 728.   ISO 11357-6, ASTM D3417.     


Temperature range

Room temperature ~ 600°C (air   cooling)

Heating rate

0.1~100°C/min

Temperature resolution

0.01°C

Temperature   0.1 ℃ accuracy

±0.01°C

Temperature   repeatability

±0.1°C

Constant   temperature time

program   setting

Temperature   control mode

PID   temperature control, temperature rise, constant temperature (automatic   program control)

Atmosphere   control gas

nitrogen,   oxygen (automatic switching of instrument)

Gas flow

0-300mL/min

Gas   pressure

0.2MPa

The   measuring range

0~±600mW

DSC   resolution

0.01μW

DSC   sensitivity

0.001μW

Display   mode

24bit color,   7 inch LCD touch screen display

Data   interface

standard USB   interface

Parameter   standard

equipped with   standard materials (indium, tin), users can correct the temperature by   themselves

Curve   scanning

temperature   rise scan, cooling scan, constant temperature scan

Display

Large LCD screen Engligh

With English software, can install on user’s computer

Power

220V, 50HZ

The instrument has   multiple sets of thermocouples, one set to test the sample temperature, and   one group to test the internal ambient temperature of the instrument.

The instrument has   multi-point correction function, which can meet the accurate testing of high,   medium and low samples

The software can set the   data acquisition frequency, and can export EXECL data packages and PDF   reports


BS EN 728

Plastics piping and ducting systems. Polyolefin pipes and fittings. Determination of oxidation induction time.


ISO 11357-6:2018
Plastics -- Differential scanning calorimetry (DSC) -- Part 6: Determination of oxidation induction time (isothermal OIT) and oxidation induction temperature (dynamic OIT)


ASTM D3417
Standard Test Method for Enthalpies of Fusion and Crystallization of Polymers by Differential Scanning Calorimetry (DSC)

      
DSC Oxidative Induction Time test sample install video
  
Differential scanning Calorimeter, Oxidative Induction Time test (OIT) operation video 1
  
Differential scanning Calorimeter, Oxidative Induction Time test (OIT) operation video 2
Your email address
Your message

Related products
contact

Email:
export@unitedtest.com
unitedtest@hotmail.com
Tel:
+86-10-57055071
+86-13911165373
SKYPE:unitedtest08
WhatsApp:
+86 13911165373

X